EasyManua.ls Logo

Endress+Hauser PROline promag 53 - Temperature Table Remote Version

Endress+Hauser PROline promag 53
20 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Promag 53 NEPSI Zone 1, Zone 21 0
Endress+Hauser 7
Temperature table remote
version
Sensor
Max. medium temperature [°C] for T1
~ T6 in relation to the maximum ambient temperature T
a
1)
with polyurethane lining;
2)
with hard rubber lining
1)
with PFA lining;
2)
with PTFE lining;
3)
versions in IP 68 are limited to 130 °C.
The minimum ambient temperature is –20 °C.
A version for ambient temperatures up to –40 °C is optionally available.
Transmitter
The transmitter of the remote version has the temperature class T6 when installing into the Ex d housing up
to an ambient temperature of T
a
= 60 °C.
The maximum ambient temperature range is –20 to +60 °C.
A version for ambient temperatures up to –40 °C is optionally available.
DN T
a
T6
(85 °C)
T5
(100 °C)
T4
(135 °C)
T3
(200 °C)
T2
(300 °C)
T1
(450 °C)
Promag 5*W**-************
25 to 1200
1)
+50 °C 50 50 50 50 50 50
50 to 2000
2)
+60 °C 80 80 80 80 80 80
DN T
a
T6
(85 °C)
T5
(100 °C)
T4
(135 °C)
T3
(200 °C)
T2
(300 °C)
T1
(450 °C)
Promag 5*P**-************
25 to 200
1)
+50 °C 80 95 130
150
3)
150
3)
150
3)
15 to 600
2)
130 130 130
25 to 200
1)
+60 °C 80 95 130 130 130 130
15 to 600
2)
DN T
a
T6
(85 °C)
T5
(100 °C)
T4
(135 °C)
T3
(200 °C)
T2
(300 °C)
T1
(450 °C)
Promag 5*H**-************
2 to 150 +50 °C 80 95 130 150 150 150
2 to 25
+60 °C
80 95 130 130 130 130
40 to 150 80 95 130 150 150 150