Promag 53 PROFIBUS DP/PA Troubleshooting
Endress+Hauser 143
473 P: PROGRESS NOTE
$
: # 473
0x80 GOOD O.K. O.K. Progress Note Cause:
End of filling process approaching.
The running filling process has exceeded the
predefined batch quantity point for the display
warning message.
Remedy:
No measures required (if necessary prepare to
replace container).
474 P: MAX. FLOW RATE
!: # 474
0x00
0x01
0x02
BAD Non specific
(uncertain
status)
O.K.
Low
High
Flow Rate Cause:
Maximum flow value entered is
exceeded.
Remedy:
Reduce the flow value.
!
Note!
Please observe Note in error message No. 471
No. Device status message
(local display)
PROFIBUS measured value status
Extended
diagnostic
message in the
PROFIBUS
master
Cause/remedy
Quality code (HEX)
Measured value status
Quality status
Quality substatus
Limits