Technical data Promag 53 PROFIBUS DP/PA
164 Endress+Hauser
Conductivity of the fluid
The minimum conductivity is 5 μS/cm (for demineralized water 20 μS/cm)
!
Note!
In the remote version, the required minimum conductivity is also influenced by the length of
the connecting cable → 19.
Limiting medium pressure range (nominal pressure)
Promag E
• EN 1092-1 (DIN 2501)
– PN 10 (DN 200 to 600 / 8 to 24")
– PN 16 (DN 65 to 600 / 3 to 24")
– PN 40 (DN 15 to 150 / ½ to 2")
• ASME B 16.5
– Class 150 (½ to 24")
• JIS B2220
– 10K (DN 50 to 300 / 2 to 12")
– 20K (DN 15 to 40 / ½ to 1½")
Promag H
The permissible nominal pressure depends on the process connection and the seal:
• 40 bar flange, weld nipple (with O-ring seal)
• 16 bar all other process connections
Promag L
• EN 1092-1 (DIN 2501)
– PN 6 (DN 350 to 2400 / 14 to 90")
– PN 10 (DN 200 to 2400 / 8 to 90")
– PN 16 (DN 25 to 300 / 1 to 12")
• EN 1092-1, lap joint flange, stampel plate
– PN 10 (DN 25 to 300 / 1 to 12")
• ASME B16.5
– Class 150 (1 to 24")
• AWWA
– Class D (28 to 90")
• AS2129
– Table E (350 to 1200 / 14 to 48")
• AS4087
– PN 16 (350 to 1200 / 14 to 48")
Promag P
• EN 1092-1 (DIN 2501)
– PN 10 (DN 200 to 600 / 8 to 24")
– PN 16 (DN 65 to 600 / 3 to 24")
– PN 25 (DN 200 to 600 / 8 to 24")
– PN 40 (DN 25 to 150 / 1 to 6")
• ASME B 16.5
– Class 150 (1 to 24")
– Class 300 (1 to 6")
• JIS B2220
– 10K (DN 50 to 300 / 2 to 12")
– 20K (DN 25 to 300 / 1 to 12")
• AS 2129
– Table E (DN 25 / 1"), 50 / 2")