EasyManua.ls Logo

Endress+Hauser PROline promag 53 - 3.3 Installation

Endress+Hauser PROline promag 53
190 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation Promag 53 PROFIBUS DP/PA
20 Endress+Hauser
3.3 Installation
3.3.1 Installing the Promag E sensor
"
Caution!
The protective covers mounted on the two sensor flanges guard the PTFE lining, which is
turned over the flanges. Consequently, do not remove these protection plates until
immediately before the sensor is installed in the pipe.
Protection plates must remain in place while the device is in storage.
Make sure that the lining is not damaged or removed from the flanges.
!
Note!
Bolts, nuts, seals, etc. are not included in the scope of supply and must be supplied by the
customer.
The sensor is designed for installation between the two piping flanges:
It is essential that you observe the necessary screw tightening torques on 20.
If grounding disks are used, follow the mounting instructions which will be enclosed with
the shipment.
A0011908
Fig. 17: Installing the Promag E sensor
Seals
Comply with the following instructions when installing seals:
PFA or PTFE lining seals are not required.
Only use seals that comply with DIN EN 1514-1 for DIN flanges.
Make sure that the seals do not protrude into the piping cross-section.
"
Caution!
Risk of short circuit! Do not use electrically conductive sealing compound such as graphite.
An electrically conductive layer could form on the inside of the measuring tube and short-
circuit the measuring signal.
Ground cable
If necessary, special ground cables can be ordered as accessories for potential equalization,
128.
For information on potential equalization and detailed installation instructions for using
ground cables, please refer to 62.
Screw tightening torques (Promag E)
Note the following points:
The tightening torques listed below are for lubricated threads only.
Always tighten the screws uniformly and in diagonally opposite sequence.
Overtightening the screws will deform the sealing faces or damage the seals.
The tightening torques listed below apply only to pipes not subjected to tensile stress.