Promag 53 PROFIBUS DP/PA Wiring
Endress+Hauser 57
4.3.3 PROFIBUS DP connection diagram
Permanent assignment board (order version 53***-***********J)
a0002308
Fig. 33: Connecting the transmitter, cable cross-section: max. 2.5 mm
(14 AWG)
A View A (field housing)
B View B (stainless steel field housing)
C View C (wall-mount housing)
a Connection compartment cover
b Cable for power supply: 85 to 260 V AC, 20 to 55 V AC, 16 to 62 V DC
Terminal No. 1: L1 for AC, L+ for DC
Terminal No. 2: N for AC, L- for DC
c Ground terminal for protective conductor
d Fieldbus cable
Terminal No. 26: B (RxD/TxD-P)
Terminal No. 27: A (RxD/TxD-N)
e Ground terminal for fieldbus cable shield
Observe the following:
– the shielding and grounding of the fieldbus cable → 49
– that the stripped and twisted lengths of cable shield to the ground terminal are as short as possible
f Service adapter for connecting service interface FXA193 (Fieldcheck, FieldCare)
g Cable for external termination:
Terminal No. 24: +5 V
Terminal No. 25: DGND