EasyManua.ls Logo

Endress+Hauser PROline promag 53 - 5.6 Hardware PROFIBUS DP Settings

Endress+Hauser PROline promag 53
190 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Promag 53 PROFIBUS DP/PA Operation
Endress+Hauser 77
5.6 Hardware PROFIBUS DP settings
5.6.1 Configuring the write protection
A jumper on the I/O board provides the means of switching hardware write protection on or
off. When the hardware write protection is switched on, it is not possible to write to the
device parameters via PROFIBUS (cyclic data transmission, e.g. via FieldCare).
#
Warning!
Risk of electric shock. Exposed components carry dangerous voltages. Make sure that the
power supply is switched off before you remove the cover of the electronics compartment.
1. Switch off power supply.
2. Remove the I/O board. 148
3. Configure the hardware write protection accordingly with the aid of the jumpers (see
Figure).
4. Installation is the reverse of the removal procedure.
a0003326
Fig. 48: Switching write protection on and off with the aid of a jumper on the I/O board
A Flexible assignment board
B Permanent assignment board
1 Jumper for switching write protection on and off
1.1 Write protection switched on = it is not possible to write to the device parameters via PROFIBUS (acyclic
data transmission, e.g. via FieldCare)
1.2 Write protection switched off (factory setting) = it is possible to write to the device parameters via PROFIBUS
(acyclic data transmission, e.g. via FieldCare).
LED Overview of LED states:
Lit continuously
Ready for operation
Not lit
Not ready for operation
Flashing
System or process error present 130
1
2
3
4
W
E
N
O
1
2
3
4
W
E
N
O
1
2
3
4
W
E
N
O
1
2
3
4
W
E
N
O
INPUT/OUTPUT 4
INPUT/OUTPUT 3
1
2
3
4
W
E
N
O
1
2
3
4
W
E
N
O
1
2
3
4
W
E
N
O
1
2
3
4
W
E
N
O
1.1
1.2
1
LED
A
1.1
1.2
1
B
LED

Table of Contents

Other manuals for Endress+Hauser PROline promag 53

Related product manuals