Promag 53 PROFIBUS DP/PA Commissioning
Endress+Hauser 89
Quick Setup "Pulsating Flow"
HOME position
F
MEASURAND
O
QUICK SETUP
N
QS PULSATING FLOW (1003)
Function No. Function name Selection with
OS
To the next function with
F
1003 QS PULSATING FLOW YES
After
F
is pressed by way of confirmation, the
Quick Setup menu calls up all the subsequent
functions in succession.
▾
Basic configuration
2002 DISPLAY DAMPING 3 s
Signal type for "CURRENT OUTPUT 1"
4004 MEASURING MODE PULS. FLOW
4005 TIME CONSTANT 3 s
Signal type for "FREQ./PULSE OUTPUT 1" (for FREQUENCY operating mode)
4206 MEASURING MODE PULS. FLOW
4208 TIME CONSTANT 0 s
Signal type for "FREQ./PULSE OUTPUT 1" (for PULSE operating mode)
4225 MEASURING MODE PULS. FLOW
Other settings
8005 ALARM DELAY 0 s
6400 ASSIGN LOW FLOW CUTOFF VOLUME FLOW
6402 ON-VALUE LOW FLOW CUTOFF Recommended setting:
a0004432-en
*Full scale values → 17
6403 OFF-VALUE LOW FLOW CUTOFF 50%
6404 PRESSURE SHOCK SUPPRESSION 0 s
▾
Back to the HOME position:
Press and hold down the
X
key for longer than three seconds or
Repeatedly press and release the
X
key Exit the function matrix step by step
1000
On-value
max. full scale (per DN)*