Proline Promag 50H, 53H
Endress+Hauser 19
Operating conditions: Process
Medium temperature range The permissible medium temperature depends on the sensor and the sealing material:
Sensor:
DN 2 to 100 (1/12 to 4"): –20 to +150 °C (–4 to +302 °F)
Seals:
EPDM: –20 to +150 °C (–4 to 302 °F)
Viton: –20 to +150 °C (–4 to 302 °F)
Silicone: –20 to +150 °C (–4 to 302 °F)
Kalrez: –20 to +150 °C (–4 to 302 °F)
Conductivity The minimum conductivity is:
5 μS/cm for fluids generally
20 μS/cm for demineralized water
!
Note!
In the remote version, the necessary minimum conductivity also depends on the cable length
( ä 17, Section "Length of connecting cable").
Medium pressure range
(nominal pressure)
The permitted nominal pressure depends on the process connection and the seal:
40 bar (580 psi):
flange EN 1092-1 (DIN 2501)
weld socket for DIN EN ISO 1127 pipes and ODT (with O-ring seal)
16 bar (232 psi): all other process connections
Pressure tightness Measuring tube lining: PFA
Limiting flow The diameter of the pipe and the flow rate determine the nominal diameter of the sensor.
The optimum flow velocity is between 2 to 3 m/s (6.5 to 9.8 ft/s). The velocity of flow (v), moreover, has to
be matched to the physical properties of the fluid:
v < 2 m/s (6.5 ft/s): for small conductivities
v > 2 m/s (6.5 ft/s): for fluids causing build-up such as high-fat milk etc.
Nominal diameter Limit values for abs. pressure [mbar] ([psi]) at fluid temperatures:
[mm] [inch] 25 °C
(77 °F)
80 °C
(176 °F)
100 °C
(212 °F)
130 °C
(266 °F)
150 °C
(302 °F)
180 °C
(356 °F)
2 to 1001/12 to 4"000000
Flow characteristic values (SI units)
Diameter Recommended flow rate Factory settings
[mm] [inch]
Min./max. full scale value
(v ~ 0.3 or 10 m/s)
Full scale value, current output
(v ~ 2.5 m/s)
Pulse value
(~ 2 pulses/s)
Low flow cut off
(v ~ 0.04 m/s)
2 1/12" 0.06 to 1.8 dm
3
/min 0.5 dm
3
/min 0.005 dm
3
0.01 dm
3
/min
41/8" 0.25 to 7dm
3
/min 2 dm
3
/min 0.025 dm
3
0.05 dm
3
/min
83/8" 1 to 30dm
3
/min 8 dm
3
/min 0.10 dm
3
0.1 dm
3
/min
15 ½" 4 to 100 dm
3
/min 25 dm
3
/min 0.20 dm
3
0.5 dm
3
/min
25 1" 9 to 300 dm
3
/min 75 dm
3
/min 0.50 dm
3
1.00 dm
3
/min
40 1½" 25 to 700 dm
3
/min 200 dm
3
/min 1.50 dm
3
3.00 dm
3
/min
50 2" 35 to 1100 dm
3
/min 300 dm
3
/min 2.50 dm
3
5.00 dm
3
/min
65 60 to 2000 dm
3
/min 500 dm
3
/min 5.00 dm
3
8.00 dm
3
/min
80 3" 90 to 3000 dm
3
/min 750 dm
3
/min 5.00 dm
3
12.0 dm
3
/min
100 4" 145 to 4700 dm
3
/min 1200 dm
3
/min 10.0 dm
3
20.0 dm
3
/min