Endress+Hauser Proline Promag L 400 - Transporting the Product

Endress+Hauser Proline Promag L 400
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Storage and transport Proline Promag L 400
12 Endress+Hauser
5.2 Transporting the product
5.2.1 For measuring devices ≤ DN 300 (12")
!
WARNING
Center of gravity of the measuring device is higher than the suspension points of the
webbing slings.
Risk of injury if the measuring device slips.
Secure the measuring device from rotating or slipping.
Observe the weight specified on the packaging (stick-on label).
Observe the transport instructions on the stick-on label on the electronics compartment
cover.
A0015606
Transport the measuring device to the measuring point in the original packaging.
Do not remove protective covers or protective caps installed on process connections.
They prevent mechanical damage to the sealing surfaces and contamination in the
measuring tube.
A0015604
5.2.2 For measuring devices ≥ DN 350 (14")
Lift the measuring device by the flange using the metal brackets.
If transporting by forklift, do not lift the sensor by the metal casing. This would buckle
the casing and damage the internal magnetic coils.