Proline Promag L 400 PROFIBUS DP Commissioning
Endress+Hauser 91
Parameter overview with brief description
Parameter Prerequsite Description Selection / User
entry
Factory setting
Assign simulation process variable – Select a process variable for
the simulation process that is
activated.
• Off
• Volume flow
• Mass flow
• Conductivity
Off
Value process variable A process variable is selected
in the Assign simulation
process variable parameter.
Enter the simulation value for
the selected process variable.
Signed floating-point
number
0
Simulation device alarm – Switch the device alarm on and
off.
• Off
• On
Off
Diagnostic event category – Select the category of the
diagnostic event.
• Sensor
• Electronics
• Configuration
• Process
Sensor
Simulation diagnostic event – Switch simulation of the
diagnostic event on and off.
For the simulation, you can
choose from the diagnostic
events of the category selected
in the Diagnostic event
category parameter.
• Off
• Picklist
Diagnostic events
(depends on the
selected category)
Off
10.9 Protecting settings from unauthorized access
The following options exist for protecting the configuration of the measuring device from
unintentional modification after commissioning:
• Write protection via access code for the local display and Web browser (→ 91)
• Write protection via write protection switch (→ 92)
• Write protection via keypad lock (→ 59)
10.9.1 Write protection via access code
The effects of the customer-specific access code are as follows:
• Via local operation, the parameters for the measuring device configuration are write-
protected and their values can no longer be changed.
• Device access via the Web browser is protected, as are the parameters for the measuring
device configuration.
Navigation
"Setup" menu → Advanced setup → Administration → Define access code
Structure of the submenu
Define access code
→
Define access code
Confirm access code
Defining the access code via local display
1. Navigate to the Enter access code parameter.
2. Define a max. 4-digit numeric code as an access code.
3. Enter the access code again to confirm the code.
The -symbol appears in front of all write-protected parameters.