Commissioning Proline Promag P 100
42 Endress+Hauser
10.2 Establishing a connection via FieldCare
•
For FieldCare connection (→ 37)
• For establishing a connection via FieldCare (→ 40)
10.3 Configuring the device address via software
Applies only to device version with EtherNet/IP communication type
Software addressing is carried out in the Communication submenu.
Navigation path
"Setup" menu → Communication
When delivered, the measuring device has the following factory settings:
Ethernet network and Web server
IP address 192.168.1.212
Subnet mask 255.255.255.0
Default gateway 192.168.1.212
• If hardware addressing is active, software addressing is disabled.
•
If a switch is changed to the "ON" position, then the first three octets of the software
address are retained.
If the IP address of the device is not known, the device address currently configured can
be read out: Operating Instructions for the device
10.4
Configuring the measuring device
The Setup menu with its submenus is used for fast commissioning of the measuring device.
The submenus contain all the parameters required for configuration, such as parameters for
measurement or communication.
Submenu Meaning
System units Configuring the units for all measured values
Medium selection Defining the medium
Communication Configuration of the digital communication interface
Low flow cut off Configuring the low flow cut off
10.5 Defining the tag name
To enable fast identification of the measuring point within the system, you can enter a unique
designation using the "Device tag" parameter and thus change the factory setting.