Proline Promag P 200
Endress+Hauser 11
HART load • Min. 250 Ω
•
Max.500 W
Dynamic variables The measured variables can be freely assigned to the dynamic variables.
Measured variables for PV (primary dynamic variable)
• Volume flow
• Mass flow
Measured variables for SV, TV, QV (secondary, tertiary and quaternary
dynamic variable)
• Volume flow
• Mass flow
• Totalizer 1
• Totalizer 2
• Totalizer 3
Power supply
Terminal assignment Transmitter
4-20 mA HART connection version with additional outputs
A0013570
A0018161
Maximum number of terminals, without integrated overvoltage
protection
Maximum number of terminals, with integrated
overvoltage protection
1
2
3
Output 1 (passive): supply voltage and signal transmission
Output 2 (passive): supply voltage and signal transmission
Ground terminal for cable shield
Order code for
"Output"
Terminal numbers
Output 1 Output 2
1 (+) 2 (-) 3 (+) 4 (-)
Option A 4-20 mA HART (passive) -
Option B
1)
4-20 mA HART (passive) Pulse/frequency/switch output
(passive)
1) Output 1 must always be used; output 2 is optional.
Supply voltage
An external power supply is required for each output. The following supply voltage values apply for the 4-20
mA HART current output:
Order code for
"Output"
Minimum terminal voltage
1)
2)
Maximum terminal voltage
• Option A: 4-20 mA HART
• Option B: 4-20 mA HART, pulse/
frequency/switch output
For 4 mA: ³ DC18 V
For 20 mA: ³ DC14 V
DC 35 V