Proline Promag P 200
12 Endress+Hauser
1)
External supply voltage of the power supply unit with load (® ä 8)
2) For device versions with local display SD03: The terminal voltage must be increased by DC 2 V if backlighting is used.
For information on the Ex connection values (® ä 9)
Various power supply units can be ordered from Endress+Hauser: see "Accessories" section (® ä 37)
Power consumption
Transmitter
Order code for
"Power supply"
Maximum power consumption
Option A: 4-20 mA HART 770 mW
Option B: 4-20 mA HART, pulse/frequency/switch output Operation with output 1:770 mW
Operation with output 1 and 2:2 770 mW
For information on the Ex connection values (® ä 9)
Current consumption
For 4-20 mA or 4-20 mA HART current output: 3.6 to 22.5 mA
If the option Defined value is selected in the Failure mode parameter (® ä 7): 3.59 to 22.5 mA
Power supply failure
Totalizers stop at the last value measured.
Configuration is retained in the device memory (HistoROM).
Error messages (incl. total operated hours) are stored.
Electrical connection Connecting the transmitter
1
2
A0015510
1
Cable entry for output 1
2 Cable entry for output 2