Proline Promag W 400
Endress+Hauser 25
2.5
[%]
2.0
1.5
1.0
0.5
0
0.2 %
0.5 %
0
10 [m/s]
v
32 [ft/s]
v
0.2
v
0.5
A0017051
 15 Flat Spec in % o.r.
Flat Spec flow values0.5 %
Nominal diameter v
0.5
[mm] [in] [m/s] [ft/s]
25 to 600 1 to 24 0.5 1.64
Flat Spec flow values0.2 %
Nominal diameter v
0.2
[mm] [in] [m/s] [ft/s]
25 to 600 1 to 24 1.5 4.92
Electrical conductivity
Max. measured error not specified.
Accuracy of outputs
o.r. = of reading; o.f.s. = of full scale value
The outputs have the following base accuracy specifications.
Current output
Accuracy Max. ±0.025 % o.f.s. or ±5 µA
Pulse/frequency output
Accuracy Max. ±50 ppm o.r.
Repeatability
o.r. = of reading
Volume flow
Max. ±0.1 % o.r. ± 0.5 mm/s (0.02 in/s)
Electrical conductivity
Max. ±5 % o.r.
Influence of ambient
temperature
o.r. = of reading; o.f.s. = of full scale value
Current output
Temperature coefficient Typically ±50 ppm/°C o.r. or ±1 µA/°C