Diagnostics and troubleshooting Proline Promag W 400 EtherNet/IP
114 Endress+Hauser
Options Description
Logbook entry only The device continues to measure. The diagnostic message is entered in the Event
logbook (events list) submenu only and is not displayed in alternation with the measured
value display.
Off The diagnostic event is ignored, and no diagnostic message is generated or entered.
12.8 Overview of diagnostic information
The amount of diagnostic information and the number of measured variables affected
increase if the measuring device has one or more application packages.
In the case of some items of diagnostic information, the status signal and the
diagnostic behavior can be changed. Adapt the diagnostic information →  113
Diagnostic
number
Short text Remedy instructions Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
Diagnostic of sensor
004 Sensor 1. Change sensor
2. Contact service
S Alarm
022 Sensor temperature 1.Change main electronic module
2.Change sensor
F Alarm
043 Sensor short circuit 1.Check sensor and cable 2.Change
sensor or cable
S Warning
062 Sensor connection 1.Check sensor connections
2.Contact service
F Alarm
082 Data storage 1. Check module connections
2. Contact service
F Alarm
083 Memory content 1. Restart device
2. Contact service
F Alarm
190 Special event 1 Contact service F Alarm
Diagnostic of electronic
201 Device failure 1. Restart device
2. Contact service
F Alarm
222 Electronic drift Change main electronic module F Alarm
242 Software incompatible 1. Check software
2. Flash or change main electronics
module
F Alarm
252 Modules incompatible 1. Check electronic modules 2.
Change electronic modules
F Alarm
262 Module connection 1. Check module connections
2. Change main electronics
F Alarm
270 Main electronic failure Change main electronic module F Alarm
271 Main electronic failure 1. Restart device
2. Change main electronic module
F Alarm
272 Main electronic failure 1. Restart device
2. Contact service
F Alarm
273 Main electronic failure Change electronic F Alarm
281 Electronic initialization Firmware update active, please
wait!
F Alarm
283 Memory content 1. Reset device 2. Contact service F Alarm