Proline Promag W 800
18 Endress+Hauser
Remote version cable
specifications
Electrode cable
3 × 0.38 mm² PVC cable with common, braided copper shield (Ø 7 mm) and individually shielded
cores
Conductor resistance: 50 km
Capacitance: core/shield: 420 pF/m
Operating temperature: –20 to +80 °C (–4 to 176 °F)
Cable cross-section: max. 2.5 mm
Coil current cable
2 × 0.75 mm² PVC cable with common, braided copper shield (Ø 7 mm)
Conductor resistance: 37 km
Capacitance: core/core, shield grounded: 120 pF/m
Operating temperature: –20 to +80 °C (–4 to 176 °F)
Cable cross-section: max. 2.5 mm²
Test voltage for cable insulation: 1433 V AC r.m.s 50/60 Hz or 2026 V DC
A0003194
Cable cross-section
a Electrode cable
b Coil current cable
1Core
2 Core insulation
3 Core shield
4 Core jacket
5 Core reinforcement
6 Cable shield
7 Outer jacket
Reinforced connecting cables
As an option, Endress+Hauser can also deliver reinforced connecting cables with an additional,
reinforcing metal braid.
Use a reinforced connecting cable in the following situations:
When laying the cable directly in the ground
Where there is a risk of damage from rodents
If using the device below IP68 degree of protection
Operation in zones of severe electrical interference
The measuring device complies with the general safety requirements in accordance with EN 61010-1
and the EMC requirements of IEC/EN 61326.
"
Caution!
Grounding is by means of the ground terminals provided for the purpose inside the connection
housing. Ensure that the stripped and twisted lengths of cable shield to the ground terminal are as
short as possible.
1
2
3
4
5
6
7
ab