Mounting procedure Proline Promag W 800
24 Endress+Hauser
2 x DN³
5 x DN³
A0041088
No inlet and outlet runs for devices with the order code for "Design": Option C.
Installation near pumps
NOTICE
Negative pressure in the measuring tube can damage the liner!
In order to maintain the system pressure, install the device in the flow direction
downstream from the pump.
Install pulsation dampers if reciprocating, diaphragm or peristaltic pumps are used.
A0041083
Information on the liner's resistance to partial vacuum →  108
Information on the measuring system's resistance to vibration and shock →  106
Installation of very heavy devices
Support required for nominal diameters of DN ≥ 350 mm (14 in).
NOTICE
Damage to the device!
If incorrect support is provided, the sensor housing could buckle and the internal magnetic
coils could be damaged.
Only provide supports at the pipe flanges.
A0041087