Mounting procedure Proline Promag W 800
24 Endress+Hauser
A0041088
No inlet and outlet runs for devices with the order code for "Design": Option C.
Installation near pumps
NOTICE
Negative pressure in the measuring tube can damage the liner!
‣
In order to maintain the system pressure, install the device in the flow direction
downstream from the pump.
‣
Install pulsation dampers if reciprocating, diaphragm or peristaltic pumps are used.
A0041083
• Information on the liner's resistance to partial vacuum → 108
• Information on the measuring system's resistance to vibration and shock → 106
Installation of very heavy devices
Support required for nominal diameters of DN ≥ 350 mm (14 in).
NOTICE
Damage to the device!
If incorrect support is provided, the sensor housing could buckle and the internal magnetic
coils could be damaged.
‣
Only provide supports at the pipe flanges.
A0041087