Mounting procedure Proline Promag W 800
28 Endress+Hauser
Installation downstream of pumps
Installation without inlet and outlet runs is possible: devices with the order code for
"Design", option C.
In the case of devices with the order code for
"Design", option J and K, an inlet run of only
≥ 2 x DN must be taken into consideration.
Installation upstream of valves
Installation without inlet and outlet runs is possible: devices with the order code for
"Design", option C.
Installation downstream of valves
The device can be installed without inlet and outlet runs if the valve is 100% open during
operation: devices with the order code for "Design", option C.
Installation dimensions
For the dimensions and installation lengths of the device, see the "Technical
Information" document, "Mechanical construction" section → 120
6.1.2 Environmental and process requirements
Ambient temperature range
Transmitter –25 to +60 °C (–13 to +140 °F)
Local display –20 to +60 °C (–4 to +140 °F); readability of the local display may be
impaired at temperatures outside the temperature range.
Sensor • Process connection material, carbon steel:
–10 to +60 °C (+14 to +140 °F)
• Process connection material, stainless steel:
–40 to +60 °C (–40 to +140 °F)
Liner Do not exceed or fall below the permitted temperature range of the liner
→ 107.
External battery pack Do not exceed or fall below the battery temperature range specified by
the manufacturer.
If operating outdoors:
• Measuring device suitable for use in wet environments
• Install the measuring device in a shady location.
• Avoid direct sunlight, particularly in warm climatic regions. Use display protection to
provide additional protection from sunlight. → 94
• Avoid direct exposure to weather conditions.