EasyManua.ls Logo

Endress+Hauser Proline Promag W 800 - Page 42

Endress+Hauser Proline Promag W 800
126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Mounting procedure Proline Promag W 800
42 Endress+Hauser
PUSH TO
REMOVE
TX 20
4 x
11.
PUSH TO
REMOVE
12.
A0044276
11. Loosen the fixing screws of the main electronics module (when reassembling, pay
attention to the tightening torque →  42).
12. Remove the main electronics module.
14.
15.
13.
4 mm
4 x
A0044277
13. Loosen the fixing screws of the transmitter housing (when reassembling, pay
attention to the tightening torque →  42).
14. Lift the transmitter housing.
15. Turn the housing to the desired position in increments of 90°.
Reassembling the transmitter housing
NOTICE
Excessive tightening torque applied to the fixing screws!
Risk of damaging the plastic transmitter.
Tighten the fixing screws as per the tightening torque: 2.5 Nm (1.8 lbf ft)
Mount the antenna on the post using the antenna bracket.
Step
→  40
Fixing screw Tightening torques
1 Housing cover 2.5 Nm (1.8 lbf ft)
9 Electronics module 0.6 Nm (0.4 lbf ft)
11 Main electronics module 1.5 Nm (1.1 lbf ft)
13 Transmitter housing 5.5 Nm (4.1 lbf ft)