Proline Promag W 800 Electrical connection
Endress+Hauser 57
Metal, ungrounded pipe
The sensor and transmitter are installed in a way that provides electrical insulation from
PE, e.g. applications for electrolytic processes or systems with cathodic protection.
Starting conditions:
Unlined metal pipe
Pipes with an electrically conductive liner
+
2.
3.
PE = P P
P M
A0042253
1. Connect the pipe flanges and transmitter via the ground cable.
2. Route the shielding of the signal lines via a capacitor (recommended value 1.5µF/
50V).
3. Device connected to power supply such that it is floating in relation to the protective
earth (isolation transformer). This measure is not required in the case of 24V DC
supply voltage without PE (= SELV power unit).
7.4 Power supply via battery packs Proline 800 -
Standard
7.4.1 Battery pack arrangement
1
A0046594
1 Battery pack