7.5.2 Inserting and connecting buffer capacitors and battery packs
The device is delivered either with the batteries already installed, or enclosed
separately, depending on national standards and guidelines. If the batteries are
inserted and connected upon delivery, to operate the device it is important to ensure
that switch "B" is set to "ON" and the ribbon cable is connected to the electronics
module.
The device starts once the buffer capacitor is connected. After 15 seconds, a measured
value appears on the display.
Connect the battery packs immediately once the buffer capacitor has been connected.
1 → 2 → 3 →
A0042838 A0043731 A0043793
‣
Open the connection
compartment cover.
‣
Open the display module.
‣
Set switch "B" to "ON".
4 → 5 →
A0044382 A0043823
‣
Remove the cover of the
battery pack.
6 → 7 →
A0043734 A0043735
‣
Insert the buffer capacitor.
‣
Plug the buffer capacitor into connector 3.