Electrical connection Proline Promag W 800
64 Endress+Hauser
2. Firmly tighten the union nut of the cable glands.
3. Pot the field housing with a potting compound.
4. Check that the housing seals are clean and fitted correctly. Dry, clean or replace the
seals if necessary.
5. Tighten all housing screws and screw covers (torque: 20 to 30 Nm).
7.9.3 Degree of protection IP66/67, type 4X enclosure, Proline 800 -
Advanced
The measuring device fulfills all the requirements for the degree of protection IP66/67,
type 4X enclosure.
To guarantee IP66/67 protection, type 4X enclosure, carry out the following steps after the
electrical connection:
1. Check that the housing seals are clean and fitted correctly. Dry, clean or replace the
seals if necessary.
2. Tighten all housing screws and screw covers.
3. Firmly tighten the cable glands.
4. To ensure that moisture does not enter the cable entry, route the cable so that it
loops down before the cable entry ("water trap").
A0029278
5. Insert dummy plugs (corresponding to the housing degree of protection) into unused
cable entries.
NOTICE
Standard dummy plugs used for transportation do not have the appropriate degree of
protection and can result in damage to the device!
Use suitable dummy plugs corresponding to the degree of protection.
7.10 Post-connection check
Are cables or the device undamaged (visual inspection)?
Do the cables used comply with the requirements?
Are the mounted cables relieved of tension?
Are all cable glands installed, securely tightened and leak-tight? Cable run with "water trap"
→  63?
Only for remote version: is the sensor connected to the right transmitter?
Check the serial number on the nameplate of the sensor and transmitter.
Does the supply voltage match the specifications on the transmitter nameplate?
Is the terminal assignment correct ?
Is the terminal assignment or the pin assignment of the device plug correct?
Batteries correctly inserted and secured?
DIP switch in correct position?