Proline Promag W 800 Maintenance
Endress+Hauser 87
Shut down device
1. Open System.
2. Open Device management.
3. Open Device reset.
4. Select Shut down device.
5. Press OK to confirm.
As soon as F418 appears on the local display, the device can be disconnected
from the power supply by means of DIP switch B without any loss of data.
If an external power supply (external battery pack) is available, the device does
not need to be switched off.
6. Replace the empty battery pack.
Replacing the empty battery pack - Promag 800
1 → 2 →
¼
1.
2.
A0046655 A0046656
Turn the cover to the right by a 1/4 turn.
Lift off the cover.
3 →
1.
2.
A0046658
Open the cover of the electronics carrier.
4 → 5 →
1. 2.
A0046666 A0046671
Take the battery out of the battery compartment (it is still attached,
however) and disconnect the battery plug.
Now remove the battery from
the battery compartment
entirely.