Proline Promag W 800 Maintenance
Endress+Hauser 89
Replacing the empty battery pack - Promag 800 Advanced
1 → 2 → 3 →
1.
2.
T20
ON
ON
1
2
1.
2.
T20
A0042838 A0043731 A0043343
Open the connection
compartment cover.
Open the display module.
Remove the cover of the
battery pack.
4 → 5 →
A0043347 A0043354
Unplug the empty battery pack.
Remove the empty battery
pack.
6 → 7 →
A0043732 A0043733
Insert the new battery pack.
Plug in the connector of the new battery pack.
The device switches back on. After 15 seconds, a measured value
appears on the display.