5.5 Software settings
5.5.1 Device address
Has to be set for measuring devices with the following communication methods:
•PROFIBUS PA
Device address range: 0 to 126, factory setting 126
The device address can be configured via:
Miniature switches see section Hardware settings 16
Local operation see description below
Note! The Commissioning Quick Setup has to be executed before setting the device address.
1.
F
Enter the function matrix
2.
P
Select the COMMUNICATION group
F
Conrm selection
3.
N
Select the BUS ADDRESS function
4.
P
Enter the desired device address
F
Conrm entry
5.
P
> 3 s = Return to the measured value display