EasyManua.ls Logo

Endress+Hauser Proline Promass 80 - Page 6

Endress+Hauser Proline Promass 80
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation Proline Promass 80
6 Endress+Hauser
The following mounting locations should be avoided:
At the highest point in a pipe (risk of air accumulating)
In an open down pipe directly upstream from a free pipe outlet. For ways of using the
measuring device in down pipes, see the related Operating Instructions on the CD-ROM.
2.2.3 Orientation
The direction of the arrow on the nameplate of the measuring device must match the flow
direction of the fluid.
The following table lists the possible orientations of the measuring devices:
Vertical Horizontal Horizontal
A0004572 A0004576 A0004576
Transmitter at the side Transmitter at the top
Transmitter at the
bottom
Promass A Recommended Possible (
m
) Possible (
m
,
p
)
Promass E Recommended Recommended (
n
) Recommended (
o
,
p
)
Promass F Recommended Recommended (
n
) Recommended (
o
,
p
)
Promass F HT*
Compact version
Recommended Not suitable Recommended (
o
,
p
)
Promass F HT*
Remote version
Recommended Possible (
n
) Recommended (
o
,
p
)
Promass H Recommended Recommended Recommended (
p
)
Promass I Recommended Recommended Recommended (
p
)
Promass P Recommended Recommended Recommended (
p
)
Promass S Recommended Recommended Recommended (
p
)
*HT = high temperature version for medium temperatures (TM) > 200 °C (> 392 °F)
m
Do not install the measuring device in such a way that is suspended without any support or securing unit.
n
This orientation is not suitable for fluids with entrained solids.
o
This orientation is not suitable for outgassing fluids.
p
This orientation is not suitable for low fluid temperatures.

Other manuals for Endress+Hauser Proline Promass 80