Installation Proline Promass 80
6 Endress+Hauser
The following mounting locations should be avoided:
• At the highest point in a pipe (risk of air accumulating)
• In an open down pipe directly upstream from a free pipe outlet. For ways of using the
measuring device in down pipes, see the related Operating Instructions on the CD-ROM.
2.2.3 Orientation
• The direction of the arrow on the nameplate of the measuring device must match the flow
direction of the fluid.
• The following table lists the possible orientations of the measuring devices:
Vertical Horizontal Horizontal
A0004572 A0004576 A0004576
Transmitter at the side Transmitter at the top
Transmitter at the
bottom
Promass A Recommended Possible (
m
) Possible (
m
,
p
)
Promass E Recommended Recommended (
n
) Recommended (
o
,
p
)
Promass F Recommended Recommended (
n
) Recommended (
o
,
p
)
Promass F HT*
Compact version
Recommended Not suitable Recommended (
o
,
p
)
Promass F HT*
Remote version
Recommended Possible (
n
) Recommended (
o
,
p
)
Promass H Recommended Recommended Recommended (
p
)
Promass I Recommended Recommended Recommended (
p
)
Promass P Recommended Recommended Recommended (
p
)
Promass S Recommended Recommended Recommended (
p
)
*HT = high temperature version for medium temperatures (TM) > 200 °C (> 392 °F)
m
Do not install the measuring device in such a way that is suspended without any support or securing unit.
n
This orientation is not suitable for fluids with entrained solids.
o
This orientation is not suitable for outgassing fluids.
p
This orientation is not suitable for low fluid temperatures.