Proline Promass 80 Installation
Endress+Hauser 7
Special installation instructions for Promass A
"
Caution!
Risk of measuring pipe fracture if sensor installed incorrectly!
The sensor may not be installed in a pipe as a freely suspended sensor:
Using the base plate, mount the sensor directly on the floor, the wall or the ceiling.
Support the sensor on a firmly mounted support base (e.g. angle bracket).
Vertical
We recommend two installation versions when mounting vertically:
Mounted directly on a wall using the base plate
Measuring device supported on an angle bracket mounted on the wall
A0018980
Horizontal
We recommend the following installation version when mounting horizontally:
Measuring device standing on a firm support base
A0018979
2.2.4 Heating
For information on the heating, please see the Operating Instructions on the CD-ROM.
2.2.5 Thermal insulation
For information on the thermal insulation, please see the Operating Instructions on the
CD-ROM.
10 mm
4 x