Device Functions Proline Promass 80 3 Group SYSTEM UNITS
Endress+Hauser 11
UNIT DENSITY Use this function to select the unit for displaying the fluid density.
The unit you select here is also valid for:
Current output
Frequency output
Switching points
Density adjustment value
Density response value for EPD
Options:
Metric
g/cm
3
; g/cc; kg/dm
3
; kg/l; kg/m
3
; SD 4 °C, SD 15 °C, SD 20 °C; SG 4 °C,
SG 15 °C, SG 20 °C
US
lb/ft
3
; lb/gal; lb/bbl (normal fluids); lb/bbl (beer); lb/bbl (petrochemicals); lb/bbl
(filling tanks)
Imperial
lb/gal; lb/bbl (beer); lb/bbl (petrochemicals)
Factory setting:
kg/l
SD = Specific Density, SG = Specific Gravity
The specific density is the ratio of fluid density to water (at water temperature = 4, 15,
20 °C)
UNIT REFERENCE
DENSITY
Use this function to select the unit for displaying the reference density.
The unit you select here is also valid for:
Current outputs
Frequency outputs
Relay switch points (limit value for density)
Fixed reference density (for calculation of corrected volume flow)
Options:
Metric:
kg/Nm
3
kg/Nl
US:
g/Scc
kg/Sm
3
lb/Scf
Factory setting:
kg/Nl
UNIT TEMPERATURE Use this function to select the unit for displaying the temperature.
The unit you select here is also valid for:
Current output
Frequency output
Switching points
Options:
°C (CELSIUS)
K (KELVIN)
°F (FAHRENHEIT)
R (RANKINE)
Factory setting:
°C (CELSIUS)