Device Functions Proline Promass 80 4 Group QUICK SETUP
Endress+Hauser 13
4 Group QUICK SETUP
!
Note!
• The display returns to the QUICK SETUP COMMISSIONING cell if you press the X key
combination during parameter interrogation.
Number reference (m – r) see next page
Function description QUICK SETUP
QUICK SETUP
COMMISSIONING
Use this function to start the Setup menu for commissioning.
Options:
NO YES
Factory setting:
NO
m The “DELIVERY SETTING” option sets every selected unit to the factory setting.
The “ACTUAL SETTING” option accepts the units you previously configured.
n Only units not yet configured in the current Setup are offered for selection in each cycle. The unit for mass,
volume and corrected volume is derived from the corresponding flow unit.
o The “YES” option remains visible until all the units have been configured. “NO” is the only option displayed when
no further units are available.
p Only outputs not yet configured in the current Setup are offered for selection in each cycle.
q The ”YES” option remains visible until all the outputs have been configured. “NO” is the only option displayed
when no further outputs are available.
r The “automatic parameterization of the display” option contains the following basic settings/
factory settings:
YES: Line 1= Mass flow; Line 2 = Totalizer 1
NO: The existing (selected) settings remain.