7 Group TOTALIZER 1/2 Device Functions Proline Promass 80
20 Endress+Hauser
7 Group TOTALIZER 1/2
Function description TOTALIZER 1/2
ASSIGN TOTALIZER Use this function to assign a measured variable to the totalizer.
Options:
MASS FLOW
VOLUME FLOW
CORRECTED VOLUME FLOW
Factory setting:
MASS FLOW
! Note!
The totalizer is reset to “0” as soon as the selection is changed.
SUM Use this function to view the total for the totalizer measured variable aggregated since
measuring commenced. The value can be positive or negative.
Display:
max. 7-digit floating-point number, including sign and unit (e.g. 15467.04 kg)
! Note!
The totalizer response to faults is defined in the FAILSAFE MODE function (see Page 22).
OVERFLOW Use this function to view the overflow for the totalizer aggregated since measuring
commenced.
Total flow quantity is represented by a floating decimal point number consisting of max.
7 digits. You can use this function to view higher numerical values (>9,999,999) as
overflows. The effective quantity is thus the total of OVERFLOW plus the value returned
by the SUM function.
Example:
Reading for 2 overflows: 2 E7 kg (= 20,000,000 kg)
The value returned by the SUM function = 196,845.7 kg
Effective total quantity = 20,196,845.7 kg
Display:
Integer with exponent, including sign and unit, e.g. 2 E7 kg