9 Group CURRENT OUTPUT 1/2 Device Functions Proline Promass 80
26 Endress+Hauser
Setting the span by means of
the 0_4 mA and 20 mA
value
(contd)
DENSITY and REFERENCE DENSITY
The density value for the 0/4 mA current is defined in the VALUE 0_4 mA function,
(input range 0.0000 to +99999). The appropriate unit is taken from the UNIT
DENSITY function.
The density value for the 20 mA current is defined in the VALUE 20 mA function,
(input range 0.0000 to +99999). The appropriate unit is taken from the UNIT
DENSITY and UNIT REF. DENSITY function.
Example (for STANDARD measuring mode):
A0004734
m = Initial value (0...20 mA)
n= Lower signal on alarm level: depends on the setting in the function CURRENT SPAN
o= Initial value (4...20 mA): depends on the setting in the function CURRENT SPAN
p = Full scale value (0/4...20 mA): depends on the setting in the function CURRENT SPAN
q = Maximum current value: depends on the setting in the function CURRENT SPAN
r = Failsafe mode (upper signal on alarm level): depends on the setting in the functions CURRENT
SPAN and FAILSAFE MODE
A = Measuring range
TEMPERATURE
The temperature value for the 0/4 mA current is defined in the VALUE 0_4 mA
function, (input range –99999 to +99999). The appropriate unit is taken from the
UNIT TEMPERATURE function.
The temperature value for the 20 mA current is defined in the VALUE 20 mA
function, (input range –99999 to +99999). The appropriate unit is taken from the
UNIT TEMPERATURE function.
! Note!
Values with different signs cannot be entered for the 0_4 mA and 20 mA values if
SYMMETRY is the option selected in the MEASURING MODE function, (see Page 54).
The message “INPUT RANGE EXCEEDED” appears on the display.
Example (for STANDARD measuring mode):
A0001223
m = Initial value (0...20 mA)
n= Lower signal on alarm level: depends on the setting in the function CURRENT SPAN
o= Initial value (4...20 mA): depends on the setting in the function CURRENT SPAN
p = Full scale value (0/4...20 mA): depends on the setting in the function CURRENT SPAN
q = Maximum current value: depends on the setting in the function CURRENT SPAN
r = Failsafe mode (upper signal on alarm level): depends on the setting in the functions CURRENT
SPAN and FAILSAFE MODE
A = Measuring range
Function description CURRENT OUTPUT 1/2
0
3
4
5
6
I[mA]
2
A
1
Q
0
3
4
5
6
I[mA]
2
A
1
Q