EasyManua.ls Logo

Endress+Hauser Proline Promass 80 - Operation; Display and operating elements

Endress+Hauser Proline Promass 80
68 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Device Functions Proline Promass 80 10 Group PULSE/FREQUENCY OUTPUT
Endress+Hauser 31
Setting the span means of the
f-min. and
f-max. value
(contd)
DENSITY and REFERENCE DENSITY
The density value for the start value frequency is defined in the VALUE F LOW
function, (input range 0.0000 to +99999). The appropriate unit is taken from the
UNIT DENSITY and UNIT REF. DENSITY function.
The density value for the end value frequency is defined in the VALUE F HIGH
function, (input range 0.0000 to +99999). The appropriate unit is taken from the
UNIT DENSITY and UNIT REF. DENSITY function.
Example (for standard measuring mode):
A0004735-en
m = Density value and reference density at which a frequency of 0 Hz should be output
(entry in VALUE F LOW function).
n = Density value and reference density at which the frequency, defined in the END
VALUE FREQ. function, should be output (entry in VALUE F HIGH function).
TEMPERATURE
The temperature value for the 0/4 mA current is defined in the VALUE F LOW
function, (input range –99999 to +99999). The appropriate unit is taken from the
UNIT TEMPERATURE function.
The temperature value for the 20 mA current is defined in the VALUE F HIGH
function, (input range –99999 to +99999). The appropriate unit is taken from the
UNIT TEMPERATURE function.
! Note!
Values with different signs cannot be entered for the VALUE F LOW and VALUE F
HIGH if SYMMETRY is the option selected in the MEASURING MODE function (see
Page 54). The message “INPUT RANGE EXCEEDED” appears on the display.
Example (for standard measuring mode):
A0004736-en
m = Temperature value at which a frequency of 0 Hz should be output (entry in VALUE
F LOW function).
n = Temperature value at which the frequency, defined in the END VALUE FREQ.
function, should be output (entry in VALUE F HIGH function).
Function description PULSE/FREQUENCY OUTPUT
125
100
Measuring range
Measured variable (amount)
Value f min
Value f max
Frequency [%]
0
125
100
Measuring range
Measured variable (amount)
Value f min
Value f ma
x
Frequency [%]
0

Other manuals for Endress+Hauser Proline Promass 80

Related product manuals