EasyManua.ls Logo

Endress+Hauser Proline Promass 80 - Group STATUS OUTPUT

Endress+Hauser Proline Promass 80
68 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Device Functions Proline Promass 80 11 Group STATUS OUTPUT
Endress+Hauser 41
11 Group STATUS OUTPUT
Function description STATUS OUTPUT
This group is not available unless the measuring device is fitted with a status output.
ASSIGN STATUS
OUTPUT
Use this function to assign a switching function to the status output.
Options:
OFF
ON (operation)
FAULT MESSAGE
NOTICE MESSAGE
FAULT MESSAGE or NOTICE MESSAGE
EMPTY PIPE DETECTION (only if function is active)
FLOW DIRECTION
LIMIT MASS FLOW
LIMIT VOLUME FLOW
LIMIT CORRECTED VOLUME FLOW
LIMIT DENSITY
LIMIT REFERENCE DENSITY
LIMIT TEMPERATURE
Factory setting:
FAULT MESSAGE
! Note!
The behavior of the status output is of the quiescent-current type, in other words the
output is closed (status output conductive) when normal, error-free measuring is in
progress.
Please read and comply with the information on the switching characteristics of the
status output (see Page 43, 44).
If you select OFF, the only function shown in this function group is this function, in
other words ASSIGN STATUS OUTPUT.
SWITCH-ON POINT
! Note!
This function is not available unless the LIMIT VALUE or FLOW DIRECTION option was
selected in the ASSIGN STATUS OUTPUT function.
Use this function to assign a value to the switch-on point (
status output conductive).
The value can be greater or less than the switch-off point. Positive and negative values
are permissible.
User input:
5-digit floating-point number
Factory setting:
0 [kg/h] or 0 [m
3
/h] or 2 [kg/l] or 200 [°C]
SWITCH-OFF POINT
! Note!
This function is not available unless the LIMIT VALUE setting was selected in the
ASSIGN STATUS OUTPUT function.
Use this function to assign a value to the switch-off point (
status output not
conductive). The value can be greater or less than the switch-on point. Positive and
negative values are permissible.
User input:
5-digit floating-point number
Factory setting:
0 [kg/h] or 0 [m
3
/h] or 2 [kg/l] or 200 [°C]

Other manuals for Endress+Hauser Proline Promass 80