PROline Promass 80 5Operation
Endress+Hauser 43
5.4.4 Device status / Error messages
You can read the extended device status, in this case, current error messages, via
Command 48. The command delivers information which are partly coded in bits (see
table below).
!
Note:
You can find a detailed explanation of the device status and error messages and their
elimination on Page 62 ff.
Byte Bit Error No. Short error description ( → Page 62 ff. )
0
0 001 Serious device error
1 011 Measuring amplifier has faulty EEPROM
2 012 Error when accessing data of the measuring amplifier EEPROM
3 not assigned –
4 not assigned –
5 not assigned –
6 not assigned –
7 not assigned –
1
0 not assigned –
1 031 S-DAT: defective or missing
2 032 S-DAT: Error accessing saved values
3 not assigned –
4 not assigned –
5 051 I/O board and the amplifier board are not compatible.
6 not assigned –
7 not assigned –
2
0 not assigned –
1 not assigned –
2 not assigned –
3 not assigned –
4 not assigned –
5 not assigned –
6 not assigned –
7 not assigned –
3
0 not assigned –
1 not assigned –
2 not assigned –
3 111 Totalizer checksum error
4 121 I/O board and the amplifier board (software versions) are not
compatible.
5 not assigned –
6 not assigned –
7 not assigned –
4 0 not assigned –