PROline Promass 80 9 Trouble-shooting
Endress+Hauser 63
S
$
SENSOR SW DAT
# 032
Sensor:
Error accessing the calibration
values stored in the S-DAT.
1. Check whether the S-DAT is
correctly plugged into the
amplifier board
→ Page 74,
76
2. Replace the S-DAT if it is
defective. Spare parts
→
Page 72
Before replacing the DAT,
check that the new,
replacement DAT is
compatible with the measuring
electronics. Check the:
– Spare part set number
– Hardware revision code
3. Replace measuring
electronics boards if
necessary.
Spare parts
→ Page 72
S
$
A / C COMPATIB.
# 051
The I/O board and the amplifier
board are not compatible.
Use only compatible modules and
boards. Check the compatibility of
the modules used.
Check the:
– Spare part set number
– Hardware revision code
S
$
CHECKSUM TOTAL
# 111
Totalizer checksum error 1. Restart the measuring device
2. Replace the amplifier board if
necessary.
Spare parts
→ Page 72
S
!
A / C COMPATIB.
# 121
Due to different software ver-
sions, I/O board and amplifier
board are only partially compati-
ble (possibly restricted function-
ality).
! Note:
– This message is only listed in
the error history.
– Nothing is displayed on the
display.
Module with lower software ver-
sion has either to be actualizied by
FieldTool with the required soft-
ware version or the module has to
be replaced.
Spare parts
→ Page 72
S
$
COMMUNIC. AMP.
# 251
Internal communication fault on
the amplifier board.
Remove the amplifier board.
Spare parts
→ Page 72
S
$
COMMUNICATION I/O
# 261
No data reception between
amplifier and I/O board or faulty
internal data transfer.
Check the BUS contacts
S
!
CURRENT RANGE
# 351
Current output 1:
Flow is out of range.
1. Change the upper or lower
limit setting, as applicable.
2. Increase or reduce flow, as
applicable.
S
!
CURRENT RANGE
# 352
Current output 2:
Flow is out of range.
1. Change the upper or lower
limit setting, as applicable.
2. Increase or reduce flow, as
applicable.
S
!
FREQUENCY RANGE
# 355
Frequency output:
Flow is out of range.
1. Change the upper or lower
limit setting, as applicable.
2. Increase or reduce flow, as
applicable.
Type Error message / No. Cause Remedy / spare part