8 Block INPUTS Device Functions Proline Promass 83 MODBUS RS485
104 Endress+Hauser
8.1.2 Function group OPERATION
INPUTS F STATUS INPUT FAA CONFIGURATION 500
OPERATION 504
Function description
INPUTS STATUS INPUT OPERATION
ACTUAL STATUS
INPUT
MODBUS register:
Data type:
Access:
5040
4305
Integer
read
Use this function to view the current level of the status input.
Display:
0 = LOW
1 = HIGH
SIMULATION
STATUS INPUT
MODBUS register:
Data type:
Access:
5041
4306
Integer
read/write
Use this function to simulate the status input, in other words to trigger the
function assigned to the status input (see the function ASSIGN STATUS INPUT
(5000) on Page 103).
Display:
0 = OFF
1 = ON
Factory setting: OFF
! Note!
The “SIMULATION STATUS INPUT” message indicates that simulation is
active.
The measuring device continues to measure while simulation is in progress, i.e.
the current measured values are output correctly via the other outputs.
" Caution!
The setting is not saved in the event of a power failure.
VALUE
SIMULATION
STATUS INPUT
MODBUS register:
Data type:
Access:
5042
4307
Integer
read/write
! Note!
The function is not visible unless the function SIMULATION STATUS INPUT
(5041) is active.
Use this function to define the level to be assumed at the status output during the
simulation. This value is used to test downstream devices and the measuring
device itself.
Options:
0 = LOW
1 = HIGH
Factory setting: LOW
" Caution!
The setting is not saved in the event of a power failure.