Proline Promass E 100
Endress+Hauser 13
Intrinsically safe values
Terminal numbers
Supply voltage Signal transmission
20 (L-) 10 (L+) 62 (A) 72 (B)
U
o
= 16.24 V
I
o
= 623 mA
P
o
= 2.45
W
With IIC
1)
: L
o
= 92.8 µH, C
o
= 0.433 μF, L
o
/R
o
= 14.6 μH/Ω
With IIB: L
o
= 372 µH, C
o
= 2.57 μF, L
o
/R
o
= 58.3 μH/Ω
For an overview and for information on the interdependencies between the gas group - sensor - nominal
diameter, see the "Safety Instructions" (XA) document for the measuring device
1) The gas group depends on the sensor and nominal diameter ff.
Transmitter
Intrinsically safe values
Order code for
"Approval"
Terminal numbers
Supply voltage Signal transmission
20 (L-) 10 (L+) 62 (A) 72 (B)
• Option BM: ATEX II2G + IECEx Z1 Ex ia, II2D Ex tb
• Option BO: ATEX II1/2G + IECEx Z0/Z1 Ex ia, II2D
•
Option BQ: ATEX II1/2G + IECEx Z0/Z1 Ex ia
• Option BU: ATEX II2G + IECEx Z1 Ex ia
• Option C2: CSA C/US IS Cl. I, II, III Div. 1
• Option 85: ATEX II2G + IECEx Z1 Ex ia + CSA C/US
IS Cl. I, II, III Div. 1
U
i
= 16.24 V
I
i
= 623 mA
P
i
= 2.45 W
L
i
= 0 µH
C
i
= 6 nF
For an overview and for information on the interdependencies between the gas group - sensor - nominal
diameter, see the "Safety Instructions" (XA) document for the measuring device
Low flow cut off
The switch points for low flow cut off are user-selectable.
Protocol-specific data HART
Manufacturer ID 0x11
Device type ID 0x4A
HART protocol revision 7
Device description files
(DTM, DD)
Information and files under:
www.endress.com
HART load Min. 250 Ω