Depending on the housing version, the transmitters can be ordered with terminals or device plugs.
Order code
"Housing"
Connection methods available
Possible options for order code
"Electrical connection"
Outputs
Power
supply
Options
A, B
Terminals Terminals Option A: coupling M20x1
Option B: thread M20x1
Option C: thread G ½"
Option D: thread NPT ½"
Options
A, B
Device plugs
→  31
Terminals Option L: plug M12x1 + thread NPT ½"
Option N: plug M12x1 + coupling M20
Option P: plug M12x1 + thread G ½"
Option U: plug M12x1 + thread M20
Options
A, B, C
Device plugs
→  31
Device plugs
→  31
Option Q: 2 x plug M12x1
Order code for "Housing":
Option A: compact, coated aluminum
Option B: compact, hygienic, stainless
Option C ultra-compact, hygienic, stainless
A0016888
 2 Terminal assignment 4-20 mA HART with pulse/frequency/switch output
1
Power supply: DC 24 V
2 Output 1: 4-20 mA HART (active)
3 Output 2: pulse/frequency/switch output (passive)
Order code
"Output"
Terminal number
Power supply Output 1 Output 2
2 (L-) 1 (L+) 27 (–) 26 (+) 25 (–) 24 (+)
Option B DC 24 V 4-20 mA HART (active) Pulse/frequency/switch
output (passive)
Order code for "Output":
Option B: 4-20 mA HART with pulse/frequency/switch output