Proline Promass E 100
Endress+Hauser 31
4-20 mA HART with pulse/frequency/switch output
Device plug for signal transmission (device side)
A0016810
Pin Assignment
1 + 4-20 mA HART (active)
2 - 4-20 mA HART (active)
3 + Pulse/frequency/switch output (passive)
4 - Pulse/frequency/switch output (passive)
5 Grounding/shielding
Coding Plug/socket
A Socket
• Recommended plug: Binder, series 763, part no. 79 3439 12 05
• When using the device in a hazardous location, use a suitably certified plug.
PROFIBUS DP
For use in the non-hazardous area and Zone 2/Div. 2.
Device plug for signal transmission (device side)
A0016811
Pin Assignment
1 Not assigned
2 A PROFIBUS DP
3 Not assigned
4 B PROFIBUS DP
5 Grounding/shielding
Coding Plug/socket
B Socket
• Recommended plug: Binder, series 763, part no. 79 4449 20 05
• When using the device in a hazardous location, use a suitably certified plug.
MODBUS RS485
Device plug for signal transmission with supply voltage (device side), MODBUS RS485 (intrinsically
safe)
A0016809
Pin Assignment
1 L+ Supply voltage, intrinsically safe
2 A
Modbus RS485 intrinsically safe
3 B
4 L- Supply voltage, intrinsically safe
5 Grounding/shielding
Coding Plug/socket
A Plug
• Recommended socket: Binder, series 763, part no. 79 3439 12 05
• When using the device in a hazardous location: Use a suitably certified socket.