Proline Promass E 100
Endress+Hauser 33
Transmitter
For device version with communication type:
• HART, PROFIBUS DP, EtherNet/IP: DC 20
to 30 V
• Modbus RS485, device version:
– For use in the non-hazardous area and Zone 2/Div. 2: DC 20 to 30 V
– For use in the intrinsically safe area: power supply via Safety Barrier Promass 100
Promass 100 safety barrier
DC 20 to 30 V
Power consumption Transmitter
Order code for "Output"
Maximum
Power consumption
Option B: 4-20 mA HART with pulse/frequency/switch output 3.5 W
Option L: PROFIBUS DP 3.5 W
Option M Modbus RS485, for use in non-hazardous areas and Zone 2/
Div. 2
3.5 W
Option M: Modbus RS485, for use in intrinsically safe areas 2.45 W
Option N: EtherNet/IP 3.5 W
Option R: PROFINET 3.5 W
Promass 100 safety barrier
Order code for "Output"
Maximum
Power consumption
Option M: Modbus RS485, for use in intrinsically safe areas 4.8 W
Current consumption Transmitter
Order code for "Output"
Maximum
Current consumption
Maximum
switch-on current
Option B: 4-20mA HART, pul./freq./switch output 145 mA 18 A (< 0.125 ms)
Option L: PROFIBUS DP 145 mA 18 A (< 0.125 ms)
Option M Modbus RS485, for use in non-hazardous
areas and Zone 2/Div. 2
90 mA 10 A (< 0.8 ms)
Option M: Modbus RS485, for use in intrinsically safe
areas
145 mA 16 A (< 0.4 ms)
Option N: EtherNet/IP 145 mA 18 A (< 0.125 ms)
Option R: PROFINET 145 mA 18 A (< 0.125 ms)
Promass 100 safety barrier
Order code for "Output"
Maximum
Current consumption
Maximum
switch-on current
Option M: Modbus RS485, for use in intrinsically safe
areas
230 mA 10 A (< 0.8 ms)
Power supply failure
• Totalizers stop at the last value measured.
• Depending on the device version, the configuration is retained in the device memory or in the
plug-in memory (HistoROM DAT).
•
Configuration is retained in the plug-in memory (HistoROM DAT).
• Error messages (incl. total operated hours) are stored.