Proline Promass E 100
34 Endress+Hauser
Electrical connection Connecting the transmitter
1 2 1 2 3 4
A B C
A0016924
A Housing version: compact, aluminum coated
B Housing version: compact hygienic, stainless
1 Cable entry or device plug for signal transmission
2 Cable entry or device plug for supply voltage
C Housing version: ultra-compact, hygienic, stainless, M12 device plug
3 Device plug for signal transmission
4 Device plug for supply voltage
Terminal assignment→  23
Pin assignment, device plug→
 30
In the case of device versions with a connector, the transmitter housing does not need to be
opened to connect the signal cable or power supply cable.
Connection examples
Current output 4 to 20 mA HART
4
4...20 mA
5
2
1
3
6
A0029055
 9 Connection example for 4 to 20 mA HART current output (active)
1
Automation system with current input (e.g. PLC)
2 Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable
specifications →  39
3 Connection for HART operating devices →
 79
4 Resistor for HART communication (≥ 250 Ω): observe maximum load
5 Analog display unit: observe maximum load
6 Transmitter