The maximum cable length for individual wire cross-sections is specified in the table below. Observe
the maximum capacitance and inductance per unit length of the cable and connection values for
hazardous areas .
Wire cross-section Maximum cable length
[mm
2
] [AWG] [m] [ft]
0.5 20 70 230
0.75 18 100 328
1.0 17 100 328
1.5 16 200 656
2.5 14 300 984
Performance characteristics
Reference operating
conditions
Error limits based on ISO 11631
Water with
+15 to +45 °C (+59 to +113 °F) at2 to 6 bar (29 to 87 psi)
Specifications as per calibration protocol
Accuracy based on accredited calibration rigs that are traced to ISO 17025.
To obtain measured errors, use the Applicator sizing tool →  88
Maximum measured error
o.r. = of reading; 1 g/cm³ = 1 kg/l; T = medium temperature
Base accuracy
Design fundamentals →  44
Mass flow and volume flow (liquids)
±0.15 % o.r.
±0.10 % o.r. (order code for "Calibration flow", option A, B, C
, for mass flow)
±0.25 % o.r.
Mass flow (gases)
±0.35 % o.r.
Density (liquids)
Under
reference operating conditions
Standard density
calibration
1)
[g/cm³] [g/cm³]
±0.0005 ±0.02
1) Valid over the entire temperature and density range
Temperature
±
0.5 °C ± 0.005
· T °C (±
0.9 °F ± 0.003 · (T – 32) °F)