The maximum cable length for individual wire cross-sections is specified in the table below. Observe
the maximum capacitance and inductance per unit length of the cable and connection values for
hazardous areas .
Wire cross-section Maximum cable length
[mm
2
] [AWG] [m] [ft]
0.5 20 70 230
0.75 18 100 328
1.0 17 100 328
1.5 16 200 656
2.5 14 300 984
Performance characteristics
Reference operating
conditions
• Error limits based on ISO 11631
• Water with
+15 to +45 °C (+59 to +113 °F) at2 to 6 bar (29 to 87 psi)
• Specifications as per calibration protocol
• Accuracy based on accredited calibration rigs that are traced to ISO 17025.
To obtain measured errors, use the Applicator sizing tool → 88
Maximum measured error
o.r. = of reading; 1 g/cm³ = 1 kg/l; T = medium temperature
Base accuracy
Design fundamentals → 44
Mass flow and volume flow (liquids)
• ±0.15 % o.r.
±0.10 % o.r. (order code for "Calibration flow", option A, B, C
, for mass flow)
• ±0.25 % o.r.
Mass flow (gases)
±0.35 % o.r.
Density (liquids)
Under
reference operating conditions
Standard density
calibration
1)
[g/cm³] [g/cm³]
±0.0005 ±0.02
1) Valid over the entire temperature and density range
Temperature
±
0.5 °C ± 0.005
· T °C (±
0.9 °F ± 0.003 · (T – 32) °F)