Proline Promass E 100
44 Endress+Hauser
[kg/m ]
3
[°C]
-40
0
-80
40 80
120 160 200 240 280 320
[°F]
-40
0
50
100 150
0
2
4
6
8
10
14
12
A0016609
20 Field density calibration, for example at +20 °C (+68 °F)
Temperature
±0.005
· T °C (± 0.005 · (T – 32) °F)
Influence of medium
pressure
The table below shows the effect on accuracy of mass flow due to a difference between calibration
pressure and process pressure.
o.r. = of reading
It is possible to compensate for the effect by:
• Reading in the current pressure measured value via the current input.
• Specifying a fixed value for the pressure in the device parameters.
Operating Instructions .
DN [% o.r./bar] [% o.r./psi]
[mm] [in]
8 ³⁄₈ no influence
15 ½ no influence
25 1 no influence
40 1½ no influence
50 2 –0.009 –0.0006
80 3 –0.020 –0.0014
Design fundamentals
o.r. = of reading, o.f.s. = of full scale value
BaseAccu = base accuracy in % o.r., BaseRepeat = base repeatability in % o.r.
MeasValue = measured value; ZeroPoint = zero point stability
Calculation of the maximum measured error as a function of the flow rate
Flow rate Maximum measured error in % o.r.
A0021332
A0021339
A0021333
A0021334