Mounting Proline Promass E TB2 HART
24 Endress+Hauser
6.3 Post-mounting check
m Is the device damaged (visual inspection)?
m
Does the measuring device conform to the measuring point specifications?
For example:
•
Process temperature (® ä 103)
• Process pressure (refer to the chapter on "Material load curves" of the "Technical Information" document)
•
Ambient temperature range (® ä 19)
•
Measuring range (® ä 97)
m
Has the correct orientation for the sensor been selected (® ä 18)?
• According to sensor type
• According to fluid temperature
• According to fluid properties (outgassing, with entrained solids)
m
Does the arrow on the sensor nameplate match the direction of flow of the fluid through the
piping(® ä 18)?
m Are the measuring point identification and labeling correct (visual inspection)?
m Is the device adequately protected from precipitation and direct sunlight?
m Are the securing screw and securing clamp tightened securely?