Commissioning Proline Promass E TB2 HART
72 Endress+Hauser
Parameters Prerequisite Description Selection/
User entry
Factory setting
Decimal places 2 A measured value is
specified in the
parameter Value 2
display.
Select the number of
digits after the decimal
point for the display
value.
• x
• x.x
• x.xx
• x.xxx
• x.xxxx
x.xx
Decimal places 3 A measured value is
specified in the
parameter Value 3
display.
Select the number of
digits after the decimal
point for the display
value.
• x
• x.x
• x.xx
• x.xxx
• x.xxxx
x.xx
Decimal places 4 A measured value is
specified in the
parameter Value 4
display.
Select the number of
digits after the decimal
point for the display
value.
• x
• x.x
• x.xx
• x.xxx
• x.xxxx
x.xx
Display interval - Set the length of time
the measured values
are displayed if the
values alternate on the
display.
1 to 10 s 5 s
Header - Select the content for
the header of the local
display.
• Device tag
• Free text
Device tag
Header text The Free text option
is selected in the
Header parameter.
Enter the text for the
header of the local
display.
Max. 12 characters
such as letters,
numbers or special
characters
- - - - - - - - - - - -
Separator - Select the decimal
separator for
numerical values.
• . (point)
• , (comma)
. (point)
10.6 Configuration management
After commissioning, you can save the current device configuration, copy it to another measuring
point or restore the previous device configuration. You can do so using the Configuration
management parameter and its options.
Navigation path
Setup ® Advanced setup ® Conf.backup disp ® Config. managem.
Functions of the parameter options
Options
Description
Execute backup A backup copy of the current device configuration in the HistoROM is saved to the display
module of the device. The backup copy comprises the transmitter data of the device.
Restore The last backup copy of the device configuration is copied from the display module to the
HistoROM of the device. The backup copy comprises the transmitter data of the device.
Duplicate The transmitter configuration from another device is duplicated to the device using the
display module.
Compare The device configuration saved in the display module is compared to the current device
configuration of the HistoROM.
Clear backup data The backup copy of the device configuration is deleted from the display module of the device.
HistoROM
A HistoROM is a "non-volatile" device memory in the form of an EEPROM.