Electrical connection Proline Promass F 100 Modbus RS485
28
7 Electrical connection
NOTICE
The measuring device does not have an internal circuit breaker.
For this reason, assign the measuring device a switch or power-circuit breaker so that
the power supply line can be easily disconnected from the mains.
Although the measuring device is equipped with a fuse, additional overcurrent
protection (maximum 16 A) should be integrated into the system installation.
7.1 Connection conditions
7.1.1 Required tools
For cable entries: Use corresponding tools
For securing clamp (on aluminum housing): Allen screw3 mm
For securing screw (for stainless steel housing): open-ended wrench 8 mm
Wire stripper
When using stranded cables: crimper for wire end ferrule
7.1.2 Requirements for connecting cable
The connecting cables provided by the customer must fulfill the following requirements.
Electrical safety
In accordance with applicable federal/national regulations.
Permitted temperature range
The installation guidelines that apply in the country of installation must be observed.
The cables must be suitable for the minimum and maximum temperatures to be
expected.
Power supply cable
Standard installation cable is sufficient.
Signal cable
Modbus RS485
The EIA/TIA-485 standard specifies two types of cable (A and B) for the bus line which
can be used for every transmission rate. Cable type A is recommended.
Cable type A
Characteristic impedance 135 to 165 Ω at a measuring frequency of 3 to 20 MHz
Cable capacitance < 30 pF/m
Wire cross-section
> 0.34 mm
2
(22 AWG)
Cable type Twisted pairs
Loop resistance ≤110 Ω/km