EasyManua.ls Logo

Endress+Hauser Proline Promass F 100 - Page 54

Endress+Hauser Proline Promass F 100
122 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commissioning Proline Promass F 100 Modbus RS485
54
Parameter overview with brief description
Parameter Description Selection Factory setting
Mass flow unit Select mass flow unit.
Result
The selected unit applies for:
Output
Low flow cut off
Simulation process variable
Unit choose list Country-specific:
kg/h
lb/min
Mass unit Select mass unit. Unit choose list Country-specific:
kg
lb
Volume flow unit Select volume flow unit.
Result
The selected unit applies for:
Output
Low flow cut off
Simulation process variable
Unit choose list Country-specific:
l/h
gal/min (us)
Volume unit Select volume unit. Unit choose list Country-specific:
l (DN > 150 (6"): m³)
gal (us)
Corrected volume flow unit Select corrected volume flow unit.
Result
The selected unit applies for:
Corrected volume flow parameter
(→  71)
Unit choose list Country-specific:
Nl/h
Sft³/min
Corrected volume unit Select corrected volume unit. Unit choose list Country-specific:
Nl
Sft³
Density unit Select density unit.
Result
The selected unit applies for:
Output
Simulation process variable
Density adjustment (Expert menu)
Unit choose list Country-specific:
kg/l
lb/ft³
Reference density unit Select reference density unit. Unit choose list Country-dependent
kg/Nl
lb/Sft³

Related product manuals