Commissioning Proline Promass F 100 Modbus RS485
60
10.4.5 Configuring the low flow cut off
The Low flow cut off submenu contains the parameters that must be set in order to
configure the low flow cut off.
Navigation
"Setup" menu → Low flow cut off
Low flow cut off
Assign process variable
→  60
On value low flow cutoff
→  60
Off value low flow cutoff
→  60
Pressure shock suppression
→  60
Parameter overview with brief description
Parameter Prerequisite Description Selection / User
entry
Factory setting
Assign process variable Select process variable for low
flow cut off.
Off
Mass flow
Volume flow
Corrected volume
flow
Mass flow
On value low flow cutoff One of the following options is
selected in the Assign process
variable parameter (→  60):
Mass flow
Volume flow
Corrected volume flow
Enter on value for low flow cut
off.
Positive floating-
point number
Depends on country
and nominal
diameter
Off value low flow cutoff One of the following options is
selected in the Assign process
variable parameter (→  60):
Mass flow
Volume flow
Corrected volume flow
Enter off value for low flow cut
off.
0 to 100.0 % 50 %
Pressure shock suppression One of the following options is
selected in the Assign process
variable parameter (→  60):
Mass flow
Volume flow
Corrected volume flow
Enter time frame for signal
suppression (= active pressure
shock suppression).
0 to 100 s 0 s