10.4.6 Configuring the partial filled pipe detection
The Partially filled pipe detection submenu contains parameters that have to be set for
configuring empty pipe detection.
Navigation
"Setup" menu → Partially filled pipe detection
‣
Partially filled pipe detection
Assign process variable
→ 61
Low value partial filled pipe detection
→ 61
High value partial filled pipe detection
→ 61
Response time part. filled pipe detect.
→ 61
Parameter overview with brief description
Parameter Prerequisite Description Selection / User
entry
Factory setting
Assign process variable – Select process variable for
partially filled pipe detection.
• Off
• Density
• Reference density
Off
Low value partial filled pipe
detection
One of the following options is
selected in the Assign process
variable parameter (→ 61):
• Density
• Reference density
Enter lower limit value for
deactivating partialy filled pipe
detection.
Signed floating-point
number
200
High value partial filled pipe
detection
One of the following options is
selected in the Assign process
variable parameter (→ 61):
• Density
• Reference density
Enter upper limit value for
deactivating partialy filled pipe
detection.
Signed floating-point
number
6 000
Response time part. filled pipe
detect.
One of the following options is
selected in the Assign process
variable parameter (→ 61):
• Density
• Reference density
Enter time before diagnostic
message is displayed for
partially filled pipe detection.
0 to 100 s 1 s