Operation Proline Promass F 100 Modbus RS485
74
Navigation
"Operation" menu → Totalizer handling
‣
Totalizer handling
Control Totalizer 1 to n
→ 74
Preset value 1 to n
→ 74
Reset all totalizers
→ 74
Parameter overview with brief description
Parameter Prerequisite Description Selection / User
entry
Factory setting
Control Totalizer One of the following options is
selected in the Assign process
variable parameter (→ 66)
of the Totalizer 1 to n
submenu:
• Volume flow
• Mass flow
• Corrected volume flow
• Target mass flow
*
• Carrier mass flow
*
Control totalizer value. • Totalize
• Reset + hold
• Preset + hold
• Reset + totalize
• Preset + totalize
Totalize
Preset value One of the following options is
selected in the Assign process
variable parameter (→ 66)
of the Totalizer 1 to n
submenu:
• Volume flow
• Mass flow
• Corrected volume flow
• Target mass flow
*
• Carrier mass flow
*
Specify start value for totalizer.
Dependency
The unit of the selected
process variable is
specified for the totalizer
depending on the
selection made in the
Assign process variable
parameter:
• Volume flow option:
Volume flow unit
parameter
• Mass flow option,
Target mass flow
option, Carrier mass
flow option: Mass
flow unit parameter
• Corrected volume
flow option:
Corrected volume
unit parameter
Signed floating-point
number
Country-specific:
• 0 kg
• 0 lb
Reset all totalizers – Reset all totalizers to 0 and
start.
• Cancel
• Reset + totalize
Cancel
* Visibility depends on order options or device settings